{"id":7603,"date":"2026-07-03T14:20:06","date_gmt":"2026-07-03T06:20:06","guid":{"rendered":"https:\/\/qualia-bio.com\/blog\/performance-qualification-for-high-containment-systems-when-real-operating-conditions-change-the-test-scope\/"},"modified":"2026-07-03T14:20:08","modified_gmt":"2026-07-03T06:20:08","slug":"performance-qualification-for-high-containment-systems-when-real-operating-conditions-change-the-test-scope","status":"publish","type":"post","link":"https:\/\/qualia-bio.com\/tr\/blog\/performance-qualification-for-high-containment-systems-when-real-operating-conditions-change-the-test-scope\/","title":{"rendered":"Y\u00fcksek G\u00fcvenlikli Sistemler i\u00e7in Performans Onaylama: Ger\u00e7ek \u00c7al\u0131\u015fma Ko\u015fullar\u0131 Test Kapsam\u0131n\u0131 De\u011fi\u015ftirdi\u011finde"},"content":{"rendered":"<p>Y\u00fcksek koruma seviyeli sistemler i\u00e7in yeterlilik planlar\u0131, genellikle projenin erken a\u015famalar\u0131nda, \u00e7o\u011fu zaman Standart Operasyon Prosed\u00fcrleri (SOP\u2019ler) tamamlanmadan ve operasyonel i\u015f ak\u0131\u015flar\u0131 net bir \u015fekilde tan\u0131mlanmadan \u00f6nce kesinle\u015ftirilir. Bu kapsam, bo\u015f veya statik test ko\u015fullar\u0131 temel al\u0131narak belirlendi\u011finde, sistem her kabul kriterini ge\u00e7se bile rutin kullan\u0131m s\u0131ras\u0131nda ar\u0131za verebilir; zira ger\u00e7ek ar\u0131za modlar\u0131, ancak operat\u00f6r\u00fcn eldivenlerin i\u00e7inden elini ge\u00e7irdi\u011fi, ge\u00e7i\u015f d\u00f6ng\u00fcleri y\u00fcklendi\u011fi veya VHP d\u00f6ng\u00fcleri ger\u00e7ek\u00e7i kap konfig\u00fcrasyonlar\u0131 \u00fczerinde \u00e7al\u0131\u015ft\u0131r\u0131ld\u0131\u011f\u0131 zaman ortaya \u00e7\u0131kar. Sonu\u00e7, devreye alma a\u015famas\u0131nda her zaman g\u00f6r\u00fcn\u00fcr olmayabilir; genellikle daha sonra proses sapmalar\u0131, tart\u0131\u015fmal\u0131 muhafaza iddialar\u0131 veya yeterlilik kan\u0131tlar\u0131n\u0131n i\u015fletim iddias\u0131n\u0131 destekleyemedi\u011fi denetim bulgular\u0131 \u015feklinde ortaya \u00e7\u0131kar. Bu makalenin destekledi\u011fi pratik yarg\u0131, ger\u00e7ek i\u015fletim ko\u015fullar\u0131n\u0131n PQ\u2019nun neleri i\u00e7ermesi gerekti\u011fini ne zaman de\u011fi\u015ftirdi\u011fini bilmek ve bunun kapsam, ikame se\u00e7imi, SOP haz\u0131rl\u0131\u011f\u0131 ve statik ko\u015fullar alt\u0131nda m\u0131 yoksa i\u015fletim ko\u015fullar\u0131 alt\u0131nda m\u0131 yeterlilik karar\u0131 verilece\u011fi konusunda ne anlama geldi\u011fini anlamakt\u0131r.<\/p>\n<h2 id=\"representativeoperatingconditionsthatchangescope\">Kapsam\u0131 de\u011fi\u015ftiren tipik \u00e7al\u0131\u015fma ko\u015fullar\u0131<\/h2>\n<p>SMEPAC testi, muhafaza cihaz\u0131n\u0131n performans\u0131n\u0131 de\u011ferlendirmek i\u00e7in kontroll\u00fc bir referans noktas\u0131 sa\u011flar; ancak ISPE k\u0131lavuzunda bunun bir laboratuvar ko\u015fullu test oldu\u011fu a\u00e7\u0131k\u00e7a belirtilmi\u015ftir. Bu test, belirli bir prosesi veya ger\u00e7ek bir \u00fcretim ortam\u0131n\u0131 temsil etmez. SMEPAC sonu\u00e7lar\u0131n\u0131 prosesi temsil eden yeterlilik kan\u0131t\u0131 olarak de\u011ferlendirmek, belirli bir sonu\u00e7 do\u011furan bir planlama hatas\u0131d\u0131r: E\u011fer bir tedarik veya kabul karar\u0131 yaln\u0131zca SMEPAC verilerine dayand\u0131r\u0131l\u0131rsa, sistem asl\u0131nda hi\u00e7bir zaman test edilmedi\u011fi bir \u00fcretim ortam\u0131 i\u00e7in kabul edilebilir.<\/p>\n<p>PQ i\u00e7in temel kapsam kriteri, ekipman\u0131n \u00fcr\u00fcn\u00fcn kurulu oldu\u011fu ve ger\u00e7ek kullan\u0131m ko\u015fullar\u0131n\u0131 yans\u0131tan \u015fartlar alt\u0131nda performans gereksinimlerini kar\u015f\u0131lay\u0131p kar\u015f\u0131lamad\u0131\u011f\u0131d\u0131r. Bu, SMEPAC\u2019\u0131n daha kat\u0131 bir yorumu de\u011fildir; tamamen farkl\u0131 bir kan\u0131t kategorisidir. Genel farmas\u00f6tik ekipman kalifikasyon ilkeleri, PQ\u2019yu, kurulu ve kalibre edilmi\u015f i\u015flev (OQ) a\u015famas\u0131n\u0131n yerini, prosesi temsil eden ko\u015fullar alt\u0131nda kan\u0131tlanm\u0131\u015f performansa b\u0131rakt\u0131\u011f\u0131 a\u015fama olarak tan\u0131mlar. Y\u00fcksek s\u0131zd\u0131rmazl\u0131k sistemlerinde bu ayr\u0131m \u00f6nemlidir; \u00e7\u00fcnk\u00fc s\u0131zd\u0131rmazl\u0131k, y\u00fck mevcut oldu\u011funda, ge\u00e7i\u015f d\u00f6ng\u00fcs\u00fc aktif oldu\u011funda veya bir operat\u00f6r eldiven portu arac\u0131l\u0131\u011f\u0131yla fiziksel olarak \u00e7al\u0131\u015ft\u0131\u011f\u0131nda farkl\u0131 davran\u0131\u015flar sergileyebilir.<\/p>\n<p>Laboratuvar baz testleri ile s\u00fcrece \u00f6zg\u00fc PQ aras\u0131ndaki yap\u0131land\u0131r\u0131lm\u0131\u015f kar\u015f\u0131tl\u0131\u011f\u0131 a\u00e7\u0131k\u00e7a belirtmek \u00f6nemlidir; zira bu ikisi, planlaman\u0131n ilk a\u015famalar\u0131nda genellikle birbirinin yerine kullan\u0131labilirmi\u015f gibi ele al\u0131nmaktad\u0131r.<\/p>\n<table>\n<thead>\n<tr>\n<th>Test T\u00fcr\u00fc<\/th>\n<th>Ko\u015fullar<\/th>\n<th>Neyi G\u00f6steriyor?<\/th>\n<th>S\u0131n\u0131rlama<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td><strong>SMEPAC (laboratuvar) testleri<\/strong><\/td>\n<td>Laboratuvar ortam\u0131, ger\u00e7ek proses malzemeleri yok<\/td>\n<td>Standart laboratuvar ko\u015fullar\u0131nda s\u0131z\u0131nt\u0131 \u00f6nleme cihaz\u0131n\u0131n performans\u0131<\/td>\n<td>Belirli bir s\u00fcreci veya ger\u00e7ek bir \u00fcretim ortam\u0131n\u0131 temsil etmez<\/td>\n<\/tr>\n<tr>\n<td><strong>Performans yeterlili\u011fi (PQ)<\/strong><\/td>\n<td>\u0130\u015fleme \u00f6zg\u00fc y\u00fck, temsili operat\u00f6rler, ger\u00e7ek veya ikame \u00fcr\u00fcn<\/td>\n<td>Ekipman, \u00fcr\u00fcn\u00fcn kurulmu\u015f ve tipik kullan\u0131m ko\u015fullar\u0131 alt\u0131nda oldu\u011fu durumlarda performans gereksinimlerini kar\u015f\u0131lar<\/td>\n<td>Standart Operasyon Prosed\u00fcrleri (SOP\u2019ler) ile koordinasyon, e\u011fitimli personel ve g\u00fcvenli vekil ko\u015fullar\u0131 gerektirir<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>Bu kar\u015f\u0131la\u015ft\u0131rmadaki s\u0131n\u0131rlama s\u00fctunu, \u00f6nemsiz bir uyar\u0131 de\u011fildir. Kapsam, standart laboratuvar ko\u015fullar\u0131ndan ger\u00e7ek proses ko\u015fullar\u0131na kayd\u0131r\u0131lmazsa, operasyonel s\u0131n\u0131rlama iddias\u0131na y\u00f6nelik herhangi bir itiraz kar\u015f\u0131s\u0131nda yeterlilik kayd\u0131n\u0131 savunmak zor olacakt\u0131r \u2014 \u00f6zellikle de dinamik kullan\u0131m\u0131n kabul edilen i\u015fletme esaslar\u0131n\u0131n bir par\u00e7as\u0131 oldu\u011fu sistemler i\u00e7in.<\/p>\n\t\t<div data-elementor-type=\"widget\" data-elementor-id=\"6744\" class=\"elementor elementor-6744\" data-elementor-post-type=\"elementor_library\">\n\t\t\t\t\t\t<div class=\"elementor-element elementor-element-6744 elementor-align-right elementor-widget elementor-widget-button\" data-id=\"6744\" data-element_type=\"widget\" data-e-type=\"widget\" data-widget_type=\"button.default\">\n\t\t\t\t<div class=\"elementor-widget-container\">\n\t\t\t\t\t\t\t\t\t<div class=\"elementor-button-wrapper\">\n\t\t\t\t\t<a class=\"elementor-button elementor-button-link elementor-size-sm\" href=\"#elementor-action%3Aaction%3Dpopup%3Aopen%26settings%3DeyJpZCI6IjI1MzMiLCJ0b2dnbGUiOmZhbHNlfQ%3D%3D\">\n\t\t\t\t\t\t<span class=\"elementor-button-content-wrapper\">\n\t\t\t\t\t\t\t\t\t<span class=\"elementor-button-text\">\u015eimdi Sorgula<\/span>\n\t\t\t\t\t<\/span>\n\t\t\t\t\t<\/a>\n\t\t\t\t<\/div>\n\t\t\t\t\t\t\t\t<\/div>\n\t\t\t\t<\/div>\n\t\t\t\t<\/div>\n\t\t\n<h2 id=\"loadpatternsandoperatorworkflowsinpqplanning\">PQ planlamas\u0131nda y\u00fckleme modelleri ve operat\u00f6r i\u015f ak\u0131\u015flar\u0131<\/h2>\n<p>Dinlenme halindeki OQ testi, sistemin kurulum an\u0131ndaki i\u015flevselli\u011fini de\u011ferlendirir. Bu test, bir operat\u00f6r\u00fcn kollar\u0131 eldiven giri\u015finden ge\u00e7en hava ak\u0131\u015f\u0131n\u0131 bozdu\u011funda, bir dizi manuel m\u00fcdahale bas\u0131n\u00e7 davran\u0131\u015f\u0131n\u0131 etkiledi\u011finde veya tekrarlanan m\u00fcdahaleler bo\u015f sistem testinde hi\u00e7 kar\u015f\u0131la\u015f\u0131lmam\u0131\u015f ergonomik kal\u0131plar ortaya \u00e7\u0131kard\u0131\u011f\u0131nda neler oldu\u011funu yans\u0131tmaz. Manuel i\u015flemlerin rutin i\u015f ak\u0131\u015f\u0131n\u0131n bir par\u00e7as\u0131 oldu\u011fu herhangi bir muhafaza sistemi i\u00e7in, OQ\u2019dan PQ\u2019ya ge\u00e7i\u015f, bir belgeleme formalitesi de\u011fil, ar\u0131za riskini azaltmaya y\u00f6nelik bir ad\u0131md\u0131r.<\/p>\n<p>Bu durum, radyofarmas\u00f6tik s\u0131cak h\u00fccrelerdeki yeterlilik testleri ile do\u011frudan benzerlik g\u00f6stermektedir; burada EANM k\u0131lavuzlar\u0131, OQ i\u00e7in durgun durum testleri ile PQ i\u00e7in \u00e7al\u0131\u015fma s\u0131ras\u0131ndaki testler aras\u0131nda a\u00e7\u0131k bir ayr\u0131m yapmaktad\u0131r \u2014 zira eldivenlerin i\u00e7inden yap\u0131lan kol hareketleri, durgun durum testleri s\u0131ras\u0131nda fark edilemeyen \u015fekillerde laminer ak\u0131\u015f d\u00fczenlerini de\u011fi\u015ftirebilir. Ayn\u0131 mant\u0131k, operat\u00f6r m\u00fcdahalelerinin normal \u00fcretimin bir par\u00e7as\u0131 oldu\u011fu her t\u00fcrl\u00fc izolat\u00f6r, eldiven kutusu veya muhafaza kab\u0131 i\u00e7in de ge\u00e7erlidir. Bu m\u00fcdahaleleri taklit etmeyen kalifikasyon testleri, i\u00e7inde kimse \u00e7al\u0131\u015fmad\u0131\u011f\u0131nda iyi performans g\u00f6steren bir sistemi kabul edebilir.<\/p>\n<p>ICH Q9(R1), bu ba\u011flamda faydal\u0131 bir risk y\u00f6netimi \u00e7er\u00e7evesi sunmaktad\u0131r: Yeterlilik testlerinin kapsam\u0131, sabit say\u0131da test \u00e7al\u0131\u015fmas\u0131na veya standartla\u015ft\u0131r\u0131lm\u0131\u015f en k\u00f6t\u00fc durum senaryosuna g\u00f6re de\u011fil, muhafaza sistemindeki bir ar\u0131zan\u0131n yol a\u00e7aca\u011f\u0131 sonu\u00e7lara ve operat\u00f6r m\u00fcdahalesinin s\u0131kl\u0131\u011f\u0131na g\u00f6re ayarlanmal\u0131d\u0131r. Manuel i\u015flemlerin s\u0131k oldu\u011fu, ergonomik k\u0131s\u0131tlamalar\u0131n bulundu\u011fu veya s\u0131ral\u0131 olarak birbirine ba\u011fl\u0131 oldu\u011fu durumlarda \u2014 \u00f6rne\u011fin BIBO filtre de\u011fi\u015fimleri, OEB4\/5 muhafaza sistemlerinde eldiven portu kullan\u0131m\u0131 veya y\u00fckl\u00fc ge\u00e7i\u015f d\u00f6ng\u00fcleri gibi \u2014 bu kal\u0131plar PQ tasar\u0131m\u0131na yans\u0131t\u0131lmal\u0131d\u0131r.<\/p>\n<table>\n<thead>\n<tr>\n<th>Test A\u015famas\u0131<\/th>\n<th>Durum<\/th>\n<th>Operat\u00f6r Kat\u0131l\u0131m\u0131<\/th>\n<th>Bekleme Durumunda Yap\u0131lan Testlerde Risk Kontrol\u00fc Konusu Ele Al\u0131nmam\u0131\u015ft\u0131r<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td><strong>OQ (bekleme durumunda)<\/strong><\/td>\n<td>Ekipman \u00e7al\u0131\u015fm\u0131yor, eldiven giymi\u015f operat\u00f6r yok<\/td>\n<td>Hi\u00e7biri<\/td>\n<td>Kol hareketinden kaynaklanan laminer ak\u0131\u015f bozulmas\u0131 ve ergonomik verimlilik \u00fczerindeki etkileri<\/td>\n<\/tr>\n<tr>\n<td><strong>PQ (\u00e7al\u0131\u015f\u0131r durumda)<\/strong><\/td>\n<td>Normal i\u015f ak\u0131\u015f\u0131, eldivenlerle yap\u0131lan manuel i\u015flemler<\/td>\n<td>Ger\u00e7ek \u00fcretimdeki gibi operat\u00f6r hareketleri<\/td>\n<td>Temsili i\u015f ak\u0131\u015f\u0131 s\u0131ras\u0131nda tespit edilen geri kalan risk alanlar\u0131<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>OQ s\u00fctununda ele al\u0131nmayan risk, ger\u00e7ek \u00fcretim hatalar\u0131n\u0131n genellikle kaynakland\u0131\u011f\u0131 noktad\u0131r. Dinlenme durumunda testten ge\u00e7en ancak temsil edici bir i\u015f ak\u0131\u015f\u0131 kapsam\u0131nda hi\u00e7 test edilmemi\u015f bir sistem, ancak bir sapma soru\u015fturmas\u0131 s\u0131ras\u0131nda veya bir denet\u00e7inin \u00e7al\u0131\u015fma s\u0131ras\u0131ndaki yeterlilik verilerini talep etmesi durumunda ortaya \u00e7\u0131kan, fark edilmeyen bir kontrol a\u00e7\u0131\u011f\u0131 yaratabilir.<\/p>\n<h2 id=\"surrogateorsafeuseconditionsforhighcontainmenttests\">Y\u00fcksek g\u00fcvenlik seviyeli testler i\u00e7in ikame veya g\u00fcvenli kullan\u0131m ko\u015fullar\u0131<\/h2>\n<p>PQ i\u00e7in temsili \u00e7al\u0131\u015fma ko\u015fullar\u0131n\u0131n tan\u0131mlanmas\u0131, hemen bir pratik k\u0131s\u0131tlama ortaya \u00e7\u0131kar\u0131r: Ger\u00e7ek API, biyolojik ajan veya aktif bile\u015fik, yeterlilik \u00e7al\u0131\u015fmas\u0131 s\u0131ras\u0131nda her zaman kullan\u0131lamaz. Testte, ger\u00e7ek ajan\u0131 yeterlilik \u00e7al\u0131\u015fmas\u0131 i\u00e7in uygunsuz k\u0131lan biyolojik, kimyasal veya \u00fcretim risklerini beraberinde getirmeden, sonu\u00e7lar\u0131n anlaml\u0131 olmas\u0131n\u0131 sa\u011flayacak kadar ger\u00e7ek malzemeyi yeterince temsil eden bir ikame madde kullan\u0131lmal\u0131d\u0131r.<\/p>\n<p>SMEPAC k\u0131lavuzu, farkl\u0131 partik\u00fcl boyutlar\u0131na ve tespit s\u0131n\u0131rlar\u0131na sahip \u00e7e\u015fitli plasebo se\u00e7eneklerini belirlemektedir. K\u0131lavuzun ayr\u0131ca a\u00e7\u0131k\u00e7a ortaya koydu\u011fu husus, plasebo se\u00e7iminin \u00f6nemidir: Tespiti kolay bir ikame madde ile test edilen bir sistem, kabul kriterlerini ge\u00e7se bile, partik\u00fcl boyutu, i\u015fleme \u00f6zellikleri veya tespit s\u0131n\u0131r\u0131 test malzemesinden farkl\u0131 olan bir bile\u015fikle kar\u015f\u0131 kar\u015f\u0131ya kald\u0131\u011f\u0131nda yine de ba\u015far\u0131s\u0131z olabilir. Bu, devreye alma a\u015famas\u0131ndan rutin \u00fcretime kadar s\u00fcrebilen gizli bir ba\u015far\u0131s\u0131zl\u0131k riskidir.<\/p>\n<p>Tedarik\u00e7iler aras\u0131nda plasebo tutarl\u0131l\u0131\u011f\u0131n\u0131n bulunmas\u0131, tedarik s\u00fcre\u00e7lerinde bu riski daha da art\u0131rmaktad\u0131r.<\/p>\n<table>\n<thead>\n<tr>\n<th>Dikkate alma<\/th>\n<th>A\u00e7\u0131kl\u0131\u011fa Kavu\u015fturulmas\u0131 Gerekenler<\/th>\n<th>Ele Al\u0131nmazsa Ortaya \u00c7\u0131kabilecek Risk<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td><strong>Placebo'nun ger\u00e7ek API ile ilgisi<\/strong><\/td>\n<td>Hangi plasebo ilac\u0131n, ger\u00e7ek aktif maddenin par\u00e7ac\u0131k boyutu, tespit s\u0131n\u0131r\u0131 ve kullan\u0131m \u00f6zelliklerine en yak\u0131n oldu\u011funu belirleyin<\/td>\n<td>Bir sistem, tespit edilmesi kolay bir plasebo ile testi ge\u00e7ebilir, ancak ger\u00e7ek bile\u015fikle kar\u015f\u0131 kar\u015f\u0131ya kald\u0131\u011f\u0131nda ba\u015far\u0131s\u0131z olabilir<\/td>\n<\/tr>\n<tr>\n<td><strong>Tedarik\u00e7iler aras\u0131nda plasebo tutarl\u0131l\u0131\u011f\u0131<\/strong><\/td>\n<td>Her bir ekipman \u00fcreticisinin SMEPAC veya yeterlilik testleri i\u00e7in ayn\u0131 plasebo kullan\u0131p kullanmad\u0131\u011f\u0131n\u0131 sorun<\/td>\n<td>Farkl\u0131 tedarik\u00e7iler aras\u0131ndaki performans kar\u015f\u0131la\u015ft\u0131rmalar\u0131 g\u00fcvenilirli\u011fini yitirerek, tedarik kararlar\u0131n\u0131 olumsuz etkiliyor<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>Tedarik s\u00fcrecinde rakip sistemlerin s\u0131z\u0131nt\u0131 \u00f6nleme performanslar\u0131n\u0131 kar\u015f\u0131la\u015ft\u0131r\u0131rken, anlaml\u0131 bir kar\u015f\u0131la\u015ft\u0131rma i\u00e7in ayn\u0131 plasebo maddesinin benzer test ko\u015fullar\u0131 alt\u0131nda kullan\u0131lmas\u0131 gerekir. Bir tedarik\u00e7i daha iri taneli, daha kolay tespit edilebilen bir ikame maddeyle test yaparken, bir di\u011feri daha ince taneli ve daha d\u00fc\u015f\u00fck tespit s\u0131n\u0131r\u0131na sahip bir malzeme kullanm\u0131\u015fsa, performans verileri do\u011frudan kar\u015f\u0131la\u015ft\u0131r\u0131lamaz; bu temelde al\u0131nan bir tedarik karar\u0131, ger\u00e7ek bile\u015fikle daha iyi performans g\u00f6sterecek sistemi de\u011fil, daha olumlu test sonucu veren sistemi se\u00e7mi\u015f olabilir.<\/p>\n<p>Herhangi bir kalifikasyon veri setini muhafaza performans\u0131n\u0131n kan\u0131t\u0131 olarak kabul etmeden \u00f6nce, hangi ikame maddenin kullan\u0131ld\u0131\u011f\u0131n\u0131, bunun sizin API\u2019nize veya ajan\u0131n\u0131za uygun olup olmad\u0131\u011f\u0131n\u0131 ve tedarik\u00e7iler aras\u0131 kar\u015f\u0131la\u015ft\u0131rmalar\u0131n tutarl\u0131 test ko\u015fullar\u0131 alt\u0131nda yap\u0131l\u0131p yap\u0131lmad\u0131\u011f\u0131n\u0131 do\u011frulay\u0131n.<\/p>\n<h2 id=\"tradeoffbetweenconfidenceandqualificationcomplexity\">G\u00fcven ile yeterlilik karma\u015f\u0131kl\u0131\u011f\u0131 aras\u0131ndaki dengeleme<\/h2>\n<p>PQ kapsam\u0131n\u0131 temsili i\u015f ak\u0131\u015flar\u0131n\u0131, y\u00fckleme ko\u015fullar\u0131n\u0131 ve operat\u00f6r\u00fcn haz\u0131r bulundu\u011fu testleri i\u00e7erecek \u015fekilde geni\u015fletmek, sistemin \u00e7al\u0131\u015fma performans\u0131na duyulan g\u00fcveni art\u0131r\u0131r. Bununla birlikte, koordinasyon y\u00fck\u00fcn\u00fc de \u00f6nemli \u00f6l\u00e7\u00fcde art\u0131r\u0131r. Temsili PQ, nihai hale getirilmi\u015f SOP\u2019ler, e\u011fitimli personel, do\u011frulanm\u0131\u015f ikame malzemeler ve savunulabilir kabul kan\u0131t\u0131 olu\u015fturacak kadar tekrarlanabilir, tan\u0131mlanm\u0131\u015f bir test dizisi gerektirir. PQ uygulamas\u0131n\u0131n genellikle planland\u0131\u011f\u0131 a\u015famada bu ko\u015fullar\u0131n hi\u00e7biri otomatik olarak mevcut de\u011fildir.<\/p>\n<p>Planlama hatas\u0131, bunu basit bir ikilem olarak ele almakt\u0131r \u2014 ya basit bir PQ y\u00fcr\u00fct\u00fclecek ya da kapsaml\u0131 bir PQ y\u00fcr\u00fct\u00fclecektir. ICH Q9(R1)\u2019deki risk temelli d\u00fc\u015f\u00fcnceyle tutarl\u0131 olan ve daha yararl\u0131 bir yakla\u015f\u0131m ise, karma\u015f\u0131kl\u0131\u011f\u0131n sonu\u00e7lara g\u00f6re ayarlanmas\u0131 gerekti\u011fidir. Rutin \u00e7al\u0131\u015fma s\u0131ras\u0131nda s\u0131z\u0131nt\u0131 \u00f6nleme sisteminin ar\u0131zalanmas\u0131 durumunda operat\u00f6r\u00fcn y\u00fcksek potensli bir bile\u015fi\u011fe veya se\u00e7ilmi\u015f bir ajana maruz kalaca\u011f\u0131 bir sistemde, koordinasyon y\u00fck\u00fc ne olursa olsun daha kapsaml\u0131 \u00e7al\u0131\u015fma s\u0131ras\u0131 testleri yap\u0131lmas\u0131 i\u00e7in g\u00fc\u00e7l\u00fc gerek\u00e7eler mevcuttur. \u0130zolasyon iddias\u0131n\u0131n yaln\u0131zca kurulu mekanik i\u015flevlere ba\u011fl\u0131 oldu\u011fu ve i\u015fletim i\u015f ak\u0131\u015f\u0131n\u0131n basit ve d\u00fc\u015f\u00fck s\u0131kl\u0131kta oldu\u011fu bir sistem i\u00e7in ise, daha s\u0131n\u0131rl\u0131 bir kapsam savunulabilir.<\/p>\n<p>Karar kriterleri keyfi de\u011fildir. Kapsam geni\u015fletmesini belirleyen \u00fc\u00e7 fakt\u00f6r vard\u0131r: operat\u00f6rlerin rutin kullan\u0131m s\u0131ras\u0131nda sistemle ne s\u0131kl\u0131kta etkile\u015fime girdi\u011fi, muhafaza alan\u0131n\u0131n bu etkile\u015fimlere ne kadar duyarl\u0131 oldu\u011fu ve yeterlilik kan\u0131tlar\u0131n\u0131n ger\u00e7ek ar\u0131za modunu kapsad\u0131\u011f\u0131n\u0131 geriye d\u00f6n\u00fck olarak kan\u0131tlaman\u0131n ne kadar zor olaca\u011f\u0131. Eldiven giri\u015fi eri\u015fiminin s\u0131k oldu\u011fu, s\u0131ral\u0131 ge\u00e7i\u015f d\u00f6ng\u00fclerine sahip veya kampanyaya g\u00f6re de\u011fi\u015fen VHP dekontaminasyon y\u00fckleri bulunan sistemler, genellikle bu de\u011fi\u015fkenli\u011fi yans\u0131tan ko\u015fullar alt\u0131nda yeterlilik testinden ge\u00e7irilmelidir. Birincil riskin statik bir bariyerin mekanik ar\u0131zas\u0131 oldu\u011fu sistemler i\u00e7in ayn\u0131 kapsam geni\u015fletmesi gerekmeyebilir.<\/p>\n<p>Kapsam\u0131n yetersiz belirlenmesinin do\u011furdu\u011fu sonu\u00e7 nettir: Bir s\u0131z\u0131nt\u0131 \u00f6nleme iddias\u0131 \u2014denetim s\u0131ras\u0131nda, bir sapma soru\u015fturmas\u0131nda veya bir maruz kalma olay\u0131n\u0131n ard\u0131ndan\u2014 sorgulan\u0131rsa, ger\u00e7ek \u00e7al\u0131\u015fma ko\u015fullar\u0131n\u0131 yans\u0131tmayan yeterlilik kan\u0131tlar\u0131n\u0131 savunmak zor olacakt\u0131r. Yeterlilik kayd\u0131, sistemin testi ge\u00e7ti\u011fini g\u00f6sterecek, ancak incelenen olay\u0131n meydana geldi\u011fi ko\u015fullar alt\u0131nda ge\u00e7ti\u011fini g\u00f6stermeyecektir.<\/p>\n\t\t<div data-elementor-type=\"widget\" data-elementor-id=\"6744\" class=\"elementor elementor-6744\" data-elementor-post-type=\"elementor_library\">\n\t\t\t\t\t\t<div class=\"elementor-element elementor-element-6744 elementor-align-right elementor-widget elementor-widget-button\" data-id=\"6744\" data-element_type=\"widget\" data-e-type=\"widget\" data-widget_type=\"button.default\">\n\t\t\t\t<div class=\"elementor-widget-container\">\n\t\t\t\t\t\t\t\t\t<div class=\"elementor-button-wrapper\">\n\t\t\t\t\t<a class=\"elementor-button elementor-button-link elementor-size-sm\" href=\"#elementor-action%3Aaction%3Dpopup%3Aopen%26settings%3DeyJpZCI6IjI1MzMiLCJ0b2dnbGUiOmZhbHNlfQ%3D%3D\">\n\t\t\t\t\t\t<span class=\"elementor-button-content-wrapper\">\n\t\t\t\t\t\t\t\t\t<span class=\"elementor-button-text\">\u015eimdi Sorgula<\/span>\n\t\t\t\t\t<\/span>\n\t\t\t\t\t<\/a>\n\t\t\t\t<\/div>\n\t\t\t\t\t\t\t\t<\/div>\n\t\t\t\t<\/div>\n\t\t\t\t<\/div>\n\t\t\n<h2 id=\"sopreadinessbeforeperformanceexecution\">Performans\u0131n ger\u00e7ekle\u015ftirilmesinden \u00f6nce SOP haz\u0131rl\u0131\u011f\u0131<\/h2>\n<p>Hen\u00fcz kesinle\u015fmi\u015f SOP\u2019larla desteklenmeyen bir sisteme y\u00f6nelik PQ uygulamas\u0131, sadece bir dok\u00fcmantasyon eksikli\u011fi de\u011fil, ayn\u0131 zamanda tekrarlanabilirlik riski de olu\u015fturur. Test dizisini uygulayan operat\u00f6r, taslak prosed\u00fcrler, gayri resmi k\u0131lavuzlar veya s\u00f6zl\u00fc talimatlara g\u00f6re \u00e7al\u0131\u015f\u0131yorsa, test, sistemin rutin \u00e7al\u0131\u015fma ko\u015fullar\u0131nda ama\u00e7land\u0131\u011f\u0131 gibi \u00e7al\u0131\u015ft\u0131\u011f\u0131n\u0131 do\u011frulayamaz; zira rutin \u00e7al\u0131\u015fma, tutarl\u0131 bir \u015fekilde izlenebilecek bir bi\u00e7imde hen\u00fcz tan\u0131mlanmam\u0131\u015ft\u0131r.<\/p>\n<p>Bu, y\u00fcksek g\u00fcvenlikli sistemlerin validasyonunda proje sonlar\u0131nda s\u0131kl\u0131kla kar\u015f\u0131la\u015f\u0131lan sorunlardan biridir. PQ genellikle devreye alma takvimine g\u00f6re planlan\u0131r ve kalifikasyon s\u00fcresi ba\u015flad\u0131\u011f\u0131nda SOP\u2019lar h\u00e2l\u00e2 inceleme a\u015famas\u0131ndad\u0131r. Sonu\u00e7 olarak, belgeler tamamlanmam\u0131\u015f halde PQ\u2019yu ger\u00e7ekle\u015ftirmek ve daha sonra SOP incelemesini tamamlamak i\u00e7in bask\u0131 olu\u015fur; bu noktada, SOP\u2019lar\u0131n son halini al\u0131rken yap\u0131lan herhangi bir prosed\u00fcr de\u011fi\u015fikli\u011fi, teknik a\u00e7\u0131dan kalifikasyon etki de\u011ferlendirmesi veya k\u0131smi yeniden kalifikasyon gerektirebilir.<\/p>\n<p>PQ uygulamas\u0131ndan \u00f6nce yap\u0131lacak pratik kontrol, ilgili SOP\u2019lerin onaylanm\u0131\u015f ve haz\u0131r oldu\u011funu, ekipman kullan\u0131m k\u0131lavuzlar\u0131n\u0131n son halini alm\u0131\u015f ve sahada bulundu\u011funu ve testi ger\u00e7ekle\u015ftirecek operat\u00f6rlerin bu prosed\u00fcrlere g\u00f6re e\u011fitim ald\u0131klar\u0131n\u0131n belgelenmi\u015f oldu\u011funu do\u011frulamay\u0131 i\u00e7ermelidir.<\/p>\n<table>\n<thead>\n<tr>\n<th>Belge\/Kay\u0131t \u00d6\u011fesi<\/th>\n<th>PQ Protokol\u00fc Neleri \u0130\u00e7ermelidir?<\/th>\n<th>Neden \u00d6nemli?<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td><strong>SOP ve k\u0131lavuz referanslar\u0131<\/strong><\/td>\n<td>\u0130lgili standart \u00e7al\u0131\u015fma prosed\u00fcrleri (SOP), kullan\u0131m k\u0131lavuzlar\u0131 ve temizlik prosed\u00fcrlerinin listesi<\/td>\n<td>Test dizilerinin, ekipman\u0131n ger\u00e7ekte nas\u0131l \u00e7al\u0131\u015ft\u0131r\u0131laca\u011f\u0131na uygun olmas\u0131n\u0131 sa\u011flar<\/td>\n<\/tr>\n<tr>\n<td><strong>Personel e\u011fitim kay\u0131tlar\u0131<\/strong><\/td>\n<td>Anahtar operat\u00f6rlerin isimleri, e\u011fitim programlar\u0131 ve belgelenmi\u015f yeterlilikleri<\/td>\n<td>PQ'yu ger\u00e7ekle\u015ftiren personelin gerekli niteliklere sahip oldu\u011funu ve sapmalar\u0131n deneyimsiz operat\u00f6rlerden kaynaklanmad\u0131\u011f\u0131n\u0131 teyit eder<\/td>\n<\/tr>\n<tr>\n<td><strong>Sorumluluklar\u0131n tan\u0131mlanmas\u0131<\/strong><\/td>\n<td>Her bir ad\u0131m\u0131 kimin ger\u00e7ekle\u015ftirece\u011fi, sonu\u00e7lar\u0131 kimin onaylayaca\u011f\u0131 ve istisnalara kimin m\u00fcdahale edece\u011fi konusunda net bir g\u00f6rev da\u011f\u0131l\u0131m\u0131<\/td>\n<td>Temsili olmayan veya tekrarlanamayan test ko\u015fullar\u0131na yol a\u00e7abilecek belirsiz devirleri \u00f6nler<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>Bu \u00e7er\u00e7evedeki sorumluluk tan\u0131mlama unsuru genellikle yeterince ayr\u0131nt\u0131l\u0131 de\u011fildir. Ara sonu\u00e7lar\u0131 kimin onaylayaca\u011f\u0131, bir test dizisini durdurma yetkisine kimin sahip oldu\u011fu veya y\u00fcr\u00fctme s\u0131ras\u0131nda sapmalar\u0131n nas\u0131l \u00fcst mercilere iletilece\u011fi net olmad\u0131\u011f\u0131nda, test ko\u015fullar\u0131, sonu\u00e7lar\u0131n tekrarlanmas\u0131n\u0131 veya savunulmas\u0131n\u0131 zorla\u015ft\u0131racak \u015fekilde sapabilir. Canl\u0131 bir PQ \u00e7al\u0131\u015fmas\u0131 s\u0131ras\u0131nda belirsiz g\u00f6rev devirleri, ekipman yeterlilik protokollerinde yer almayan bir tekrarlanabilirlik ba\u015far\u0131s\u0131zl\u0131\u011f\u0131 riskidir; ancak protokol g\u00f6zden ge\u00e7irildi\u011finde denetim bulgular\u0131nda ortaya \u00e7\u0131kar.<\/p>\n<p>FDA s\u00fcre\u00e7 do\u011frulama k\u0131lavuzu, belgelenmi\u015f prosed\u00fcrlerin ve nitelikli personelin, anlaml\u0131 bir yeterlilik de\u011ferlendirmesinin ger\u00e7ekle\u015ftirilmesi i\u00e7in \u00f6n ko\u015fullar oldu\u011fu, bunlar\u0131n ayn\u0131 anda tamamlanabilecek paralel faaliyetler olmad\u0131\u011f\u0131 ilkesini desteklemektedir. SOP\u2019nin haz\u0131rl\u0131\u011f\u0131n\u0131, sonradan tamamlanacak bir belgeleme \u00e7\u0131kt\u0131s\u0131 olarak de\u011fil, uygulama \u00f6ncesi bir inceleme a\u015famas\u0131 olarak ele almak, yeterlilik kayd\u0131n\u0131 koruyan tutumdur.<\/p>\n<h2 id=\"pqtriggerfornonstaticcontainmentclaims\">Statik olmayan s\u0131n\u0131rlama iddialar\u0131 i\u00e7in PQ tetikleyicisi<\/h2>\n<p>Her muhafaza sistemi, tam \u00e7al\u0131\u015fma ko\u015fullar\u0131nda PQ testine tabi tutulmay\u0131 gerektirmez. Bu gereklili\u011fin ortaya \u00e7\u0131kmas\u0131, yaln\u0131zca ekipman kategorisi veya yasal s\u0131n\u0131fland\u0131rmaya ba\u011fl\u0131 de\u011fildir; sistemin muhafaza iddias\u0131n\u0131n, durgun durum testleriyle do\u011frulanamayan dinamik ko\u015fullara ba\u011fl\u0131 olup olmad\u0131\u011f\u0131na g\u00f6re belirlenir.<\/p>\n<p>Manuel i\u015flemlerin rutin kullan\u0131m\u0131n bir par\u00e7as\u0131 oldu\u011fu sistemlerde \u2014 bir izolat\u00f6rdeki eldiven portu eri\u015fimi, kanall\u0131 bir muhafaza muhafazas\u0131nda BIBO filtre de\u011fi\u015fimleri, bir BSL-3 ge\u00e7i\u015f odas\u0131ndaki operat\u00f6re ba\u011fl\u0131 transfer dizileri \u2014 muhafaza iddias\u0131 sabit de\u011fildir. Bu ko\u015fullar alt\u0131ndaki hava ak\u0131\u015f d\u00fczenleri, bas\u0131n\u00e7 kademeleri ve bariyer b\u00fct\u00fcnl\u00fc\u011f\u00fc, sistemin at\u0131l durumundaki davran\u0131\u015f\u0131ndan farkl\u0131l\u0131k g\u00f6sterebilir. \u0130zolasyon performans\u0131, operat\u00f6r\u00fcn ne yapt\u0131\u011f\u0131na ve sistemin bu m\u00fcdahalelere nas\u0131l tepki verdi\u011fine ba\u011fl\u0131 oldu\u011funda, OQ kan\u0131tlar\u0131 tek ba\u015f\u0131na bu iddiay\u0131 desteklemez. Radyofarmas\u00f6tik kalifikasyon uygulamalar\u0131ndan al\u0131nan s\u0131cak h\u00fccre benzetmesi bunu do\u011frudan g\u00f6stermektedir: K\u0131lavuz, \u00f6zellikle eldiven i\u00e7inden yap\u0131lan kol hareketlerinin, bo\u015fta test s\u0131ras\u0131nda g\u00f6r\u00fcnmeyen \u015fekillerde izolasyon alan\u0131n\u0131 de\u011fi\u015ftirebilece\u011fi i\u00e7in, \u00e7al\u0131\u015fmama durumundaki OQ ile \u00e7al\u0131\u015fma durumundaki PQ aras\u0131nda ayr\u0131m yapmaktad\u0131r.<\/p>\n<p>Statik olmayan iddialar i\u00e7in \u00e7al\u0131\u015fma s\u0131ras\u0131nda PQ\u2019nun bulunmamas\u0131, sadece eksik bir test seti de\u011fildir; bu, do\u011frulanmam\u0131\u015f bir iddiad\u0131r. Sistemin \u00e7al\u0131\u015fma esaslar\u0131, rutin operat\u00f6r m\u00fcdahaleleri s\u0131ras\u0131nda s\u0131z\u0131nt\u0131 \u00f6nlemenin s\u00fcrd\u00fcr\u00fcld\u00fc\u011f\u00fcn\u00fc varsay\u0131yorsa ve bu ko\u015fullar alt\u0131nda hi\u00e7bir zaman yeterlilik testi yap\u0131lmam\u0131\u015fsa, s\u00f6z konusu iddia kan\u0131tlanmam\u0131\u015ft\u0131r. Bu eksiklik, bir s\u00fcre\u00e7 sapmas\u0131 operat\u00f6r\u00fcn haz\u0131r bulundu\u011fu bir olay\u0131 i\u00e7erdi\u011finde veya bir denet\u00e7i, aktif kullan\u0131m s\u0131ras\u0131nda s\u0131z\u0131nt\u0131 \u00f6nleme derecesini destekleyen kan\u0131tlar\u0131n neler oldu\u011funu sordu\u011funda \u00f6nemli bir sorun haline gelir.<\/p>\n<p>Uygulamaya y\u00f6nelik tetikleme kriteri olduk\u00e7a a\u00e7\u0131kt\u0131r: Bir operat\u00f6r\u00fcn sistemde bulunup sistemi \u00e7al\u0131\u015ft\u0131rd\u0131\u011f\u0131 durumlarda, sistemin at\u0131l durumda oldu\u011fu durumlara k\u0131yasla kabul iddias\u0131 farkl\u0131 veya potansiyel olarak daha k\u00f6t\u00fc oluyorsa, PQ bu iddiay\u0131 desteklemek i\u00e7in \u00e7al\u0131\u015fma ko\u015fullar\u0131n\u0131 da i\u00e7ermelidir. Entegre edilmi\u015f muhafaza sistemleri i\u00e7in <a href=\"https:\/\/qualia-bio.com\/tr\/product\/module-laboratory\/bsl-3-bsl-4-module-laboratory\/\">BSL-3\/BSL-4 mod\u00fcl laboratuvarlar\u0131<\/a> operat\u00f6r i\u015f ak\u0131\u015flar\u0131n\u0131n s\u00fcrekli ve \u00e7e\u015fitlilik g\u00f6sterdi\u011fi durumlarda ya da <a href=\"https:\/\/qualia-bio.com\/tr\/product\/spacevhp\/vhp-hydrogen-peroxide-generator-type-i\/\">VHP transfer ve dekontaminasyon sistemleri<\/a> D\u00f6ng\u00fc performans\u0131 y\u00fck yap\u0131land\u0131rmas\u0131na ba\u011fl\u0131 oldu\u011funda, tetikleme ko\u015fulu neredeyse her zaman mevcuttur. Proje planlama a\u015famas\u0131nda bu tetikleyici ne kadar erken tespit edilirse, PQ y\u00fcr\u00fctmesi ba\u015flad\u0131\u011f\u0131nda kapsam geni\u015fletmesinin yol a\u00e7aca\u011f\u0131 aksakl\u0131klar o kadar az olur.<\/p>\n<p>Dekontaminasyon yeterlilik kapsam\u0131n\u0131 belirleyen ekipler i\u00e7in, <a href=\"https:\/\/qualia-bio.com\/tr\/blog\/vhp-validation-protocol-iq-oq-pq-for-hydrogen-peroxide-systems\/\">Hidrojen peroksit sistemleri i\u00e7in IQ, OQ ve PQ\u2019yu kapsayan VHP do\u011frulama protokol\u00fc<\/a> yeterlilik a\u015famalar\u0131 boyunca faz s\u0131n\u0131rlar\u0131n\u0131n ve kabul kriterlerinin genellikle nas\u0131l yap\u0131land\u0131r\u0131ld\u0131\u011f\u0131na dair yararl\u0131 bir referans sunar.<\/p>\n<p>Y\u00fcksek g\u00fcvenlikli bir ortamda en \u00f6nemli yeterlilik kan\u0131t\u0131, sistemin kontroll\u00fc, bo\u015f veya at\u0131l ko\u015fullarda testlerden ge\u00e7ti\u011fine dair kan\u0131t de\u011fildir; as\u0131l \u00f6nemli olan, sistemin rutin kullan\u0131m s\u0131ras\u0131nda fiilen mevcut olacak ko\u015fullar alt\u0131nda performans g\u00f6sterdi\u011fine dair kan\u0131tt\u0131r. PQ kapsam\u0131n\u0131 kesinle\u015ftirmeden \u00f6nce yap\u0131lmas\u0131 gereken en \u00f6nemli do\u011frulamalar \u015funlard\u0131r: kabul iddias\u0131n\u0131n dinamik mi yoksa statik mi oldu\u011fu, hangi ikame maddenin sistemin i\u00e7erece\u011fi ger\u00e7ek malzemeyi en yak\u0131ndan temsil etti\u011fi ve ilk test \u00e7al\u0131\u015ft\u0131rmas\u0131 ba\u015flamadan \u00f6nce SOP\u2019lerin ve e\u011fitimli personelin haz\u0131r olup olmad\u0131\u011f\u0131. Bu \u00fc\u00e7 kontrol, operasyonel ko\u015fullar alt\u0131nda herhangi bir itirazla kar\u015f\u0131la\u015f\u0131lmas\u0131 durumunda, olu\u015fturulan yeterlilik kayd\u0131n\u0131n muhafaza iddias\u0131n\u0131 destekleyip destekleyemeyece\u011fini belirleyecektir.<\/p>\n<p>Daha az g\u00f6ze \u00e7arpan bir risk ise ikame madde se\u00e7imi. Kaba, kolayca tespit edilebilen bir plasebo ile ba\u015far\u0131l\u0131 olan ancak ger\u00e7ek API\u2019nin partik\u00fcl boyutu veya i\u015fleme \u00f6zelliklerini yans\u0131tan bir maddeyle hi\u00e7bir zaman test edilmemi\u015f bir yeterlilik de\u011ferlendirmesi, g\u00f6r\u00fcn\u00fc\u015fte kan\u0131tlad\u0131\u011f\u0131 \u015feyi asl\u0131nda kan\u0131tlamam\u0131\u015ft\u0131r. Kar\u015f\u0131la\u015ft\u0131rmal\u0131 yeterlilik verilerini kabul etmeden \u00f6nce plasebo uygunlu\u011funu ve tedarik\u00e7inin tutarl\u0131l\u0131\u011f\u0131n\u0131 do\u011frulamak, hem yeterlilik kayd\u0131n\u0131 hem de bunu \u00f6nceleyen tedarik karar\u0131n\u0131 koruyan bir tedarik ve do\u011frulama inceleme ad\u0131m\u0131d\u0131r.<\/p>\n<h2 id=\"frequentlyaskedquestions\">S\u0131k\u00e7a Sorulan Sorular<\/h2>\n<p><strong>S: Y\u00fcksek g\u00fcvenlikli sistemimiz tamamen otomatiktir ve rutin manuel eldiven giri\u015fi veya operat\u00f6r m\u00fcdahalesi gerektirmez. \u00c7al\u0131\u015fma s\u0131ras\u0131nda PQ yap\u0131lmas\u0131na ili\u015fkin tavsiye h\u00e2l\u00e2 ge\u00e7erli midir?<\/strong><br \/>\nC: Mutlaka \u00f6yle de\u011fil. E\u011fer s\u0131z\u0131nt\u0131 \u00f6nleme iddias\u0131 yaln\u0131zca statik bariyer b\u00fct\u00fcnl\u00fc\u011f\u00fcne ve operat\u00f6r\u00fcn varl\u0131\u011f\u0131ndan etkilenmeyen otomatik s\u00fcre\u00e7lere dayan\u0131yorsa, devre d\u0131\u015f\u0131 durumdaki OQ yeterli olabilir. Ancak yine de, de\u011fi\u015fken y\u00fck konfig\u00fcrasyonlar\u0131, ge\u00e7i\u015f d\u00f6ng\u00fcleri veya farkl\u0131 tank y\u00fckleriyle yap\u0131lan VHP dekontaminasyonu gibi di\u011fer dinamik ko\u015fullar\u0131n, performans\u0131 \u00e7al\u0131\u015fma s\u0131ras\u0131nda test yap\u0131lmas\u0131n\u0131 gerektirecek \u015fekilde de\u011fi\u015ftirmedi\u011fini do\u011frulamal\u0131s\u0131n\u0131z.<\/p>\n<p><strong>S: PQ kapsam\u0131m\u0131z\u0131n, temsili operat\u00f6r i\u015f ak\u0131\u015flar\u0131n\u0131 ve y\u00fckleme ko\u015fullar\u0131n\u0131 i\u00e7ermesi gerekti\u011fine karar verdik. Testlerin y\u00fcr\u00fct\u00fclmesine ba\u015flamadan \u00f6nce atmam\u0131z gereken bir sonraki ad\u0131m nedir?<\/strong><br \/>\nC: \u0130lgili t\u00fcm Standart \u00c7al\u0131\u015fma Prosed\u00fcrlerinin (SOP) nihai hale getirilip onayland\u0131\u011f\u0131ndan ve testleri ger\u00e7ekle\u015ftirecek operat\u00f6rlerin bu prosed\u00fcrlere g\u00f6re resmi olarak e\u011fitildi\u011finden emin olun. Taslak veya gayri resmi k\u0131lavuzlara g\u00f6re PQ\u2019nun y\u00fcr\u00fct\u00fclmesi, tekrarlanabilirlik riski yarat\u0131r; testlerden sonra SOP\u2019lerde de\u011fi\u015fiklik olmas\u0131 durumunda, yeterlilik kan\u0131tlar\u0131 art\u0131k operasyonel iddiay\u0131 desteklemeyebilir.<\/p>\n<p><strong>S: D\u00fczenleyici kurumlar\u0131n, belirli bir i\u015f ak\u0131\u015f\u0131ndan ba\u011f\u0131ms\u0131z olarak, i\u015fletim s\u0131ras\u0131nda PQ\u2019yu otomatik olarak zorunlu k\u0131ld\u0131\u011f\u0131 belirli bir biyog\u00fcvenlik seviyesi veya OEB aral\u0131\u011f\u0131 var m\u0131?<\/strong><br \/>\nC: Evrensel bir e\u015fik de\u011feri bulunmamaktad\u0131r. D\u00fczenleyici beklentiler, sabit bir s\u0131n\u0131fland\u0131rmaya de\u011fil, riske dayal\u0131d\u0131r. Bununla birlikte, s\u0131z\u0131nt\u0131 durumunda ciddi sonu\u00e7lar do\u011furabilecek BSL-3\/4 veya y\u00fcksek OEB sistemlerinde, hem risk hem de operat\u00f6r etkile\u015fimi do\u011fas\u0131 gere\u011fi y\u00fcksek oldu\u011fundan, genellikle sa\u011flam bir i\u015fletme s\u0131ras\u0131nda PQ beklenir. Her durumda, ICH Q9(R1) k\u0131lavuzu, test kapsam\u0131n\u0131n ger\u00e7ek kullan\u0131m s\u0131ras\u0131nda ar\u0131za olas\u0131l\u0131\u011f\u0131 ve sonu\u00e7lar\u0131na g\u00f6re ayarlanmas\u0131n\u0131 desteklemektedir.<\/p>\n<p><strong>S: Ger\u00e7ek bile\u015fi\u011fimizi en iyi \u015fekilde temsil eden bir ikame maddenin se\u00e7ilmesi ile, PQ s\u0131ras\u0131nda kontaminasyon veya g\u00fcvenlik riskinin ortaya \u00e7\u0131kmas\u0131n\u0131 \u00f6nleme gereklili\u011fi aras\u0131nda nas\u0131l bir denge kurabiliriz?<\/strong><br \/>\nC: Par\u00e7ac\u0131k boyutu, kullan\u0131m \u00f6zellikleri ve tespit s\u0131n\u0131r\u0131 a\u00e7\u0131s\u0131ndan ger\u00e7ek API\u2019niz veya ajan\u0131n\u0131zla uyumlu olan en g\u00fcvenli ikame maddesini se\u00e7in. SMEPAC k\u0131lavuzunda \u00e7e\u015fitli plasebo se\u00e7enekleri sunulmaktad\u0131r; se\u00e7iminizin bilimsel gerek\u00e7esini belgelendirin. Tedarik\u00e7ileri kar\u015f\u0131la\u015ft\u0131r\u0131yorsan\u0131z, hepsinin kar\u015f\u0131la\u015ft\u0131r\u0131labilir test ko\u015fullar\u0131 alt\u0131nda ayn\u0131 ikame maddesini kulland\u0131\u011f\u0131ndan emin olun; aksi takdirde performans verileri anlaml\u0131 bir \u015fekilde kar\u015f\u0131la\u015ft\u0131r\u0131lamaz.<\/p>\n<p><strong>S: Temsili, \u00e7al\u0131\u015f\u0131r durumdaki PQ i\u00e7in harcanan ekstra koordinasyon ve zaman her zaman hakl\u0131 m\u0131d\u0131r, yoksa iyi bir \u015fekilde y\u00fcr\u00fct\u00fclen durgun durumdaki OQ de denetimleri ayn\u0131 derecede ba\u015far\u0131yla ge\u00e7ebilir mi?<\/strong><br \/>\nC: Bu her zaman hakl\u0131 bir gerek\u00e7e de\u011fildir. E\u011fer muhafaza iddian\u0131z esasen statikse \u2014 \u00f6rne\u011fin, operat\u00f6rlerin rutin eri\u015fiminin olmad\u0131\u011f\u0131, kal\u0131c\u0131 olarak s\u0131zd\u0131rmaz hale getirilmi\u015f bir bariyer gibi \u2014 ve operat\u00f6r m\u00fcdahaleleri nadir ve d\u00fc\u015f\u00fck riskliyse, belgelenmi\u015f bir risk gerek\u00e7esine dayanan \u201cdinlenme halinde\u201d OQ savunulabilir olabilir. Ancak, operat\u00f6r i\u015f ak\u0131\u015flar\u0131, y\u00fck de\u011fi\u015fkenli\u011fi veya s\u0131k m\u00fcdahaleler muhafaza performans\u0131n\u0131 belirliyorsa, kan\u0131tlanmam\u0131\u015f bir iddian\u0131n getirece\u011fi denetim ve g\u00fcvenlik risklerine k\u0131yasla, \u201c\u00e7al\u0131\u015fma s\u0131ras\u0131nda\u201d PQ\u2019nun maliyeti \u00f6nemsiz kal\u0131r.<\/p>","protected":false},"excerpt":{"rendered":"<p>PQ\u2019yu yaln\u0131zca temsili y\u00fckler, operat\u00f6rler veya proses ko\u015fullar\u0131n\u0131n muhafaza veya dekontaminasyon performans\u0131n\u0131 etkileyebilece\u011fi durumlarda kullan\u0131n.<\/p>","protected":false},"author":1,"featured_media":0,"comment_status":"closed","ping_status":"","sticky":false,"template":"","format":"standard","meta":{"site-sidebar-layout":"default","site-content-layout":"","ast-site-content-layout":"default","site-content-style":"default","site-sidebar-style":"default","ast-global-header-display":"","ast-banner-title-visibility":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"","ast-breadcrumbs-content":"","ast-featured-img":"","footer-sml-layout":"","ast-disable-related-posts":"","theme-transparent-header-meta":"","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","astra-migrate-meta-layouts":"default","ast-page-background-enabled":"default","ast-page-background-meta":{"desktop":{"background-color":"var(--ast-global-color-4)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"tablet":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"mobile":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""}},"ast-content-background-meta":{"desktop":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"tablet":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"mobile":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""}},"n8n_meta_description":"Use PQ only where representative loads, operators or process conditions can change containment or decontamination performance.","footnotes":""},"categories":[21],"tags":[],"class_list":["post-7603","post","type-post","status-publish","format-standard","hentry","category-blog"],"_links":{"self":[{"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/posts\/7603","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/comments?post=7603"}],"version-history":[{"count":1,"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/posts\/7603\/revisions"}],"predecessor-version":[{"id":7604,"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/posts\/7603\/revisions\/7604"}],"wp:attachment":[{"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/media?parent=7603"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/categories?post=7603"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/qualia-bio.com\/tr\/wp-json\/wp\/v2\/tags?post=7603"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}